Classic Series Vacuums Fuse Test and Repair

WARNING

For continued protection against risk of fire, replace the fuse with a 1/3 amp 250V Slo-Blo type only.

The fuse protects the low voltage control system against overload failure, mainly short circuits. It is located in the electrical box and accessed on top of the electrical box on single models and the side of the twin models.

Classic Series Vacuums Fuse Test and Repair Classic Series Vacuums Fuse Test and Repair

Fuse Check

1.Turn power off.

Classic Series Vacuums Fuse

2.Remove fuse from electrical box. Push in and turn cap counter clockwise.

Classic Series Vacuums Fuse

3.Test continuity. Set meter to Ω. Place meter probes on both fuse terminals.

4.Reinstall into electrical box. Push in and turn cap clockwise until it locks in place.

Meter Reading Status Required Action
OL - off line Thumbs Down Replace Fuse - 1/3 amp 250V Slo-Blo fuse
Continuity checks ok. Visually looks ok. Thumbs Up Fuse is Good