Classic Series Vacuums Inlet Strainer Test and Repair
The vacuum inlet strainer keeps solid waste from going into the pump from the operatory line. Empty and clean the inlet strainer everyday. Make sure gasket is seated in the groove of the filter bowl and the wire mesh strainer is straight and not collapsed.
Clean Vacuum Inlet Strainer
If screen collapses, it must be replaced.
1.Turn power off.
2.Remove filter bowl, screen, and o-ring.
3.Clean filter bowl and screen.
4.Seat o-ring into groove.
5.Install filter bowl.
6.Turn power on.