224 Exam Chair Receptacles Test and Repair

224 Exam Chair Table Receptacles Test and Repair

This illustration shows only the fuses and connections that affect the Receptacles.

Problem Required Action
No power at receptacles. Check if receptacle circuit breaker(s) tripped. Refer to: User Guide (-001 thru -003 or -011)
Test fuses at chair base.
Check output voltage from isolation transformer (115 VAC).