626 Chair Position Calibration Procedure Test and Repair

626 Chair Locate the PCB 626 Chair Position Calibration

626 Chair Calibrat Start 626 Chair Calibrat position 626 Chair calibrat complete

1.Power up chair.

2.Locate the Diagnostic PC Board.

3.Press and release the Up or Down button until Calibrat Position is displayed.

4.Press Enter button once. The display shows Calibrat Start?.

5.Press Enter button again. The chair base and back starts to cycle up and down. Calibrating (Scrolling) Position is displayed.

6.You will hear a steady pattern of beeps. When calibration is completed you will hear three short beeps and the Calibrat Complete is displayed.