626 Chair Drawer Heater / Fuses and Connections Test and Repair
This illustration shows only the fuses and connections that affect the Drawer Heater.
Problem | Required Action |
---|---|
Heater does not warm up. | Test fuses at chair base. |
Test Drawer Heater Switch. | |
Test Drawer Heater Plate. | |
Check output voltage from Isolation Transformer. |

Unplug power cord before removing receptacle cover.
1.Remove receptacle cover.
2.Remove two screws and place receptacle aside.
3.Disconnect the white wires from heater switch to the heater plate.
4.Plug in power cord.
5.Set meter to 200 VAC.
6.Depress switch ON.
7.Place meter probes on two white wires from drawer heater switch.
Meter Reading | Required Action |
---|---|
0 VAC | Replace switch. |
Approximately 115 VAC | Switch OK Refer to: Drawer Heater Plate |
Unplug power cord before re-connecting electrical leads.
8.Re-connect electrical leads.

Unplug power cord before removing receptacle cover.
1.Remove receptacle cover.
2.Remove two screws and place receptacle aside.
3.Disconnect the white wires from heater switch to the heater plate.
4.Set meter to Ω.
5.Place meter probes on the two white wires from heater plate.
Meter Reading | Required Action |
---|---|
0 Ohms | Replace Heater Plate. |
Approximately 500 Ohms | Heater Plate OK |
Unplug power cord before re-connecting electrical leads.
6.Re-connect electrical leads.

Unplug power cord before removing receptacle cover.
1.Remove receptacle cover.
2.Remove two screws and place receptacle aside.
3.Disconnect the black and red wires from isolation transformer to heater switch.
4.Plug in power cord.
5.Set meter to 200 VAC.
6.Place meter probes on the black and red wires from isolation transformer.
Meter Reading | Required Action |
---|---|
0 VAC | Replace Transformer. |
Approximately 115 VAC | Transformer OK. Refer to: Drawer Heater Switch. |
7.Re-connect electrical leads.