626 Chair Receptacles / Fuses and Connections Test and Repair

Model |
626 -001 and 626-002 |
Serial Number | All |
This illustration shows only the fuses and connections that affect the Receptacles.
Problem | Required Action |
---|---|
No power at receptacles. | Check if receptacle circuit breaker(s) tripped. Refer to: User guide. |
Test fuses at chair base. | |
Check output voltage from isolation transformer. |

Model |
626 -003 thru -006 |
Serial Number | All |
This illustration shows only the fuses and connections that affect the Receptacles.
Problem | Required Action |
---|---|
No power at receptacles. | Check if receptacle circuit breaker(s) tripped. Refer to: User guide. |
Test fuses at chair base. | |
Check output voltage from isolation transformer. |