630 Chair Base Actuator / Limit Switches Test and Repair
This illustration shows the base limit switches and the serviceable components of the base actuator. Use the table below to isolate the malfunction.
Anytime the limit switch wires or the actuator wires are disconnected, the PC Board must be calibrated.
Anytime an Overtravel Limit Switch is detected open, both Up and Down functions are deactivated.
Problem | Required Action |
---|---|
Motor runs, but makes grinding noise | Clean / lube actuator threads. Lube threads w/lithium grease. Replace actuator if necessary |
Motor runs, but chair does not move | Inspect / replace motor coupler |
Motor does not run | Test Limit Switch(es) |