641 Chair Main System Transformer Test and Repair
This transformer affects all table functions except table receptacles.
1.Set meter to 200 VAC.
2.Place meter probes on black and white wires of transformer connector. Check meter reading.
3.Place meter probes on red and orange wires of transformer connector. Check meter reading.
Black and White Wires
Meter Reading |
Status | Required Action |
---|---|---|
42 to 58 VAC |
|
Transformer OK |
Less than 42 VAC more than 58 VAC |
|
Perform Resistance Test |
Red and Orange Wires
Meter Reading |
Status | Required Action |
---|---|---|
23 to 43 VAC |
|
Transformer OK |
Less than 23 VAC |
|
Perform Resistance Test |
1.Unplug table.
2.Disconnect four transformer wires from appliance inlet.
3.Set meter to 200 Ω.
4.Place meter probes on blue/white and brown/white wires. Check meter reading.
5.Place meter probes on blue and brown wires. Check meter reading.
Meter Reading | Status | Required Action |
---|---|---|
Less than 10 Ω |
|
Transformer OK |
OL or more than 10 Ω |
|
Replace transformer. |