641 Chair Specifications
Equipment not suitable for use in the presence of a flammable anesthetic mixture with air, or with oxygen, or nitrous oxide.
For packaged weight, contact Midmark Customer Service
Specifications | |
---|---|
Patient Weight - maximum |
450 lbs - 204 kgs |
Range of Motion |
See illustration |
Dimensions |
See illustration |
Weight of Table: |
|
w/ Standard Base |
440 lbs - 200 kg |
w/ Rotational Base |
525 lbs - 238 kg |
Power Cord |
8 ft. - 243.8 cm long |
Electrical Requirements: |
See Compliance Charts |
Foot / Hand Control Voltage |
10 VAC, SELV - Safety Extra Low Voltage |
Duplex Receptacle - optional maximum load |
115 VAC, 5A, 50/60 Hz |
Fuses: |
|
Main Power - located at power cord inlet |
|
F1 and F2 |
6.3A, 250V, Type-T, 5 x 20 mm High-Breaking Current |
Duty Cycle |
|
Actuator motor run time: |
Intermittent Operation - 30 seconds On / 5 minutes off |
Classifications |
Class 1, Type B Applied Part, |
Protection against ingress of fluids |
Ordinary Equipment |
Foot Control only |
PX1 |
Regulatory Compliance: |
See Compliance Charts |
Table Speeds - by function |
|
Back / Tilt: |
|
w/Patient Load <200lbs - <90 kg |
less than 16 seconds |
w/Patient Load >200lbs - >90 kg) |
less than 20 seconds |
Base |
less than 20 seconds |
Foot |
less than 14 seconds |