230 Universal Procedures ChairWeights, Dimensions and Electrical Specifications

Equipment not suitable for use in the presence of a flammable anesthetic mixture with air, or with oxygen, or nitrous oxide.

230 and 630 Chair Weights, Dimensions and Electrical Specifications 230 and 630 Chair Weights, Dimensions and Electrical Specifications

230 and 630 Chair Weights, Dimensions and Electrical Specifications 230 and 630 Chair Weights, Dimensions and Electrical Specifications

Weights, Dimensions, Electrical Specifications

Maximum Patient Weight

450 lbs (204.1 kgs)

Maximum size Paper Roll

18 in. long x 3.5 in. diameter (45.7 x 8.9 cm)

Range of Motion and Dimensions

See illustration

Weight of 230 - w/standard base

455 lbs (206.4 kg)

Weight of 230 - w/standard base, w/packaging and skid

510 lbs (231.3 kg)

Weight of 230 - w/rotational base

540 lbs (244.9 kg)

Weight of 230 - w/rotational base, w/packaging and skid

600 lbs (271.1 kg)

Power Cord

8 ft. (243.8 cm) long

Electrical Requirements

See Compliance Charts

Foot / Hand Control Voltage

10 VAC, SELV (Safety Extra Low Voltage)

Duplex Receptacle (optional) maximum load

115 VAC, 3A, 50/60 Hz

Main Power Fuse (located at power cord inlet): F1 and F2

6.3A, 250V, Type-T, 5 x 20 mm High-Breaking Current

Upholstery Heater Fuse (located on uph. heater PC board): F1 and F2

1/2 A, 250V, Type-T, 1-1/4" x 1/4"

Duty Cycle - motor run time

Intermittent Operation. 30 seconds On / 5 minutes off.

Classifications

Class 1, Type B Applied Part

Protection against ingress of fluids

Ordinary Equipment

Foot Control only

IPX1

Regulatory Compliance:

See Compliance Charts

Chair Speeds by function

Base / Back / Tilt (all models)

15 - 25 seconds

Foot (230)

less than 20 seconds