230 Universal Procedures Chair Isolation Transformer / Chair Receptacles Output Voltage Test and Repair
Model |
230 -002 |
Serial Number | All |
This transformer affects chair receptacles only.
1.Set meter to 200 VAC .
2.Place meter probes on receptacle screws on top and bottom. Check meter reading.
Meter Reading |
Status |
Required Action |
---|---|---|
120 (+12) VAC |
|
Transformer OK. Replace receptacle(s). |
Less than 108 VAC |
|
If chair power light is On, replace transformer. If not, check supply voltage. |