230 Universal Procedures Chair Isolation Transformer / Chair Receptacles Output Voltage Test and Repair

Model

230 -002
Serial Number All

This transformer affects chair receptacles only.

230 and 630 Chair Output Voltage Test

1.Set meter to 200 VAC .

2.Place meter probes on receptacle screws on top and bottom. Check meter reading.

Meter Reading

Status

Required Action

120 (+12) VAC

Thumbs up

Transformer OK. Replace receptacle(s).

Less than 108 VAC

Thumbs down

If chair power light is On, replace transformer. If not, check supply voltage.