230 Universal Procedures Chair Main System Transformer Test and Repair

This transformer affects all chair functions except chair receptacles.
1.Set meter to 200 VAC.
2.Place meter probes on black and white wires of transformer connector. Check meter reading.
3.Place meter probes on red and orange wires of transformer connector. Check meter reading.
Black and White Wires Meter Reading |
Status |
Required Action |
---|---|---|
42 to 58 VAC |
|
Transformer OK |
< 42 VAC > 58 VAC |
|
Perform Resistance Test |
Red and Orange Wires Meter Reading |
Status |
Required Action |
---|---|---|
23 to 43 VAC |
|
Transformer OK |
< 23 VAC |
|
Perform Resistance Test |

1.Unplug chair.
2.Disconnect four transformer wires from appliance inlet.
3.Set meter to 200 Ω
4.Place meter probes on blue/white and brown/white wires. Check meter reading.
5.Place meter probes on blue and brown wires. Check meter reading.
Meter Reading |
Status |
Required Action |
---|---|---|
Less than 10 Ω |
|
Transformer OK |
OL or more than 10 Ω |
|
Replace transformer |