230 Universal Procedures Chair Primary Fuses Test and Repair
Voltage setting must match facility supply voltage. Incorrect setting will result in damage to the chair.

If Power Indicator Light is off there is no power to chair. Test the fuses.
If Power Indicator Light is on the chair has power and the fuses are OK

1.Set meter to 2 ohms.
2.Place meter probes on each end of fuse.
Meter Reading |
Status |
Required Action |
---|---|---|
OL |
|
Replace both Fuses |
Less than 5 Ω |
|
Fuse Ok |