230 Universal Procedures Chair Rotational Base Brake System Electrical Test / Fuse Location Test and Repair

Model

230-003
Serial Number All

230 and 630 Chair Rotational Base Brake System Electrical Test / Fuse Location

1.Remove PC board cover. Refer to: PC Board Cover

2.Set meter to VAC.

230 and 630 Chair Rotational Base Brake System Electrical Test / Fuse Location

3.Place meter probes on wires at J2 on Rotational Base PC Board.

4.Check voltage. If no voltage is detected, press and release the brake pedal. Check voltage again.

Meter Reading

Status

Required Action

15 - 21 VAC

Thumbs up

PC board is OK. Perform Electro-magnet Test.

0 VAC

Thumbs down

Check Rot. Base PC board fuse. If fuse is OK, replace PC board.