230 Universal Procedures Chair Rotational Base Brake System Electro-magnet Test and Repair
Model |
230-003 |
Serial Number | All |
1.Remove PC board cover. Refer to: PC Board Cover
2.Set meter to Ω
3.Disconnect electro-magnet wire harness from J2 on Rotational Base PC Board.
4.Place meter probes on wire harness.
Meter Reading |
Status |
Required Action |
---|---|---|
approximately 10 Ω |
|
Electro-magnet is OK. Check mechanical components. |
OL or less than 5 Ω |
|
Inspect magnet wire harness. If OK, replace electro-magnet. |