Ultra Chair SW1 Switch Settings for Motion Programming Test and Repair
Mode |
Switch |
Setting |
Description |
---|---|---|---|
Normal Mode |
|
1-off 2-off |
In Normal Mode all directional buttons function normally. Programmable position buttons, 1 thru 4, can be programmed for four different positions. |
Cuspidor Return |
|
1-off 2-On |
Cuspidor Position Return Mode allows operator to program a stored position on button 4 for a patient cuspidor position. Pressing button 4 not only moves chair into position for patient to use cuspidor but also activates cuspidor solenoid output on chair PCB. Output, (J13 on p.c. board), can be attached to an optional cuspidor solenoid for automatic bowl flush. |
Last Position Return |
|
1-On 2-off |
Last Position Return Mode allows user to move between two Remembered positions. A Remembered Position is a position stored by chair software anytime chair has set motionless for at least 1/2 second. Pressing button 4 will move chair to last remembered position. Pressing button 4 again toggles the chair between the last two Remembered positions |
Calibration Procedure The purpose of Calibration Mode is to determine the end of travel potentiometer sensors settings, store them in the PCB software, and use this stored information to keep any movement slightly inside these settings or limits. Calibration is an automatic routine. During calibration the chair completes two cycles. The first cycle finds the extreme end of travel sensor settings. The second cycle checks to assure the settings have been successfully stored and that they can be recalled by the software |
|||
Calibration Mode
|
|
1-On 2-On |
Remove power from chair, place both SW1 dip switches in the On position, apply power to chair, depress SW2 Calibration switch to initiate the Calibration Cycle. When complete remove chair power, place both SW1 dip switches in the off position.
|
Calibration Switch |
|
Press to initiate Calibration Procedure |
Operator Controls Calibration Mode
Program keys, 1 thru 4 on Operator Controls, or Direction keys do not work when in Calibration Mode. The chair can be moved using Manual Override by pressing the program (P) key and the desired Directional key at the same time.