Ultra Chair Specifications
Description |
Data |
---|---|
Maximum PatientWeight |
300 lbs (136 kg) |
Weight of a Unit Without Shipping Carton: |
|
Chair Only |
290 lbs (132 kg) |
Chair (LR or Console) |
310 lbs (141 kg) |
Weight of a Unit With Shipping Carton: |
|
Chair Only |
350 lbs (159 kg) |
Chair (LR or Console) |
370 lbs (168 kg) |
Hydraulic Fluid Requirements: |
|
System Capacity |
1 Quart (950 ml) |
Type of Hydraulic Fluid |
ISO VG 32 |
Electrical Requirements: |
|
115 VAC +/-10%, 5A, 60 HZ, single phase | |
|
or 230 VAC +/-10%, 2.5A, 50/60 HZ, single phase |
Fuse Ratings - Fuses are located on the PC Board: |
|
F1 Line Fuse, 115 VAC |
T6.3 AL, 250 VAC |
F1 Line Fuse, 230 VAC |
T3.15 AL, 250 VAC |
F2 Line Fuse, 115 VAC |
T6.3 AL, 250 VAC |
F2 Line Fuse, 230 VAC |
T3.15 AL, 250 VAC |
F3 Transformer Fuse |
115 VAC / 230 VAC, T0.125 AL / T0.08 AL, 250 VAC, 5 x 20mm, Type Slo-Blo |
Duty Cycle |
Intermittent Operation - 30 seconds on / 5 minutes off, motor run time |
Should it be necessary to transport the chair, its position is critical to ensure safety. Before attempting to move / transport chair, it must be lowered to its minimum height and locked in place at 0° rotation. Midmark® strongly recommends disassembling any accessory arms (unit, light, etc.) from chair before attempting to move / transport an entire operatory.
Dimensions