Ultra Chair Manual Override of Chair Positioning Test and Repair

Caution

This procedure is to be performed by the operator only if a chair malfunctions, with a patient still on chair in a raised or reclined position, to allow for safe patient exit from chair. After patient exits from the chair, discontinue use of chair until chair has been properly repaired.

Operation

1.Using touchpad, Item 1, or foot control, Item 2, depress and hold the PROGRAM button (A) and then depress desired manual positioning button(s) (B) CHAIR Back Down, CHAIR Back Up, CHAIR SEAT Down, or CHAIR Back Up until desired position is achieved.

Manual Override

Manual Override