Asepsis 21® Delivery and International (Whip) Delivery Troubleshooting
Use these troubleshooting tables to help you determine the cause of a delivery system malfunction.


Symptom | Probable Cause | Check | Correction |
---|---|---|---|
There is no water or air present at any of the handpieces or syringe(s).
|
Air on/off valve is turned off or malfunctioning. |
Check that air on/off valve is turned on. |
Contact licensed plumber to replace air on/off valve. |
Master on/off switch valve is turned off or malfunctioning. |
Check that master on/off switch valve is turned on. |
Replace master on/off switch valve if necessary. |

Symptom | Probable Cause | Check | Correction |
---|---|---|---|
None of the handpieces will run, but coolant water and syringe water work properly. |
Main air regulator turned down too low or malfunctioning. |
Check that air regulator output pressure is 80 PSI. |
Adjust air regulator output pressure until it is 80 PSI. |
Drive air foot pedal valve in foot control is malfunctioning or stuck closed. |
Check for proper air flow out of foot control when foot control pedal is depressed. |
Repair damaged foot control. If necessary, replace foot control. |
|
Pilot operated valve block is malfunctioning or has dirt in it. |
Clean valve block or replace suspect pilot operated valve block with known working valve block. |
Clean pilot operated valve block. Replace, if necessary. |
|
Air tubing from pilot operated valve block to foot control to delivery head is damaged, loose, or disconnected. |
Check air tubes for damage, looseness, or disconnection. |
Replace damaged tubing. Reconnect any loose or disconnected tubing. |
|
Syringe air and water work properly, but handpieces do not run or have water. |
Air tubing from air regulator to foot control damaged, loose, or disconnected. |
Check air tube for damage, looseness, or disconnection. |
Replace damaged tubing. Reconnect any loose or disconnected tubing. |
Handpieces have water but do not run. |
Drive air foot pedal valve is stuck closed. |
Replace suspect foot control with known working foot control. |
Clean out drive air foot pedal valve. If necessary, replace entire foot control -foot control switch valves are not available. |
Tubing for drive air from foot pedal valve to delivery head is damaged, loose, or disconnected. |
Check for air leaking from tubing and check air tubing for damage, looseness, or disconnection |
Replace damaged tubing. Reconnect any loose or disconnected tubing. |

Symptom | Probable Cause | Check | Correction |
---|---|---|---|
One handpiece does not run or have water. All other handpieces work properly. |
Handpiece on/off valve is malfunctioning or stuck open keeping kink valve “kinked”. |
Replace suspect handpiece on/off valve with known working on/off valve. |
Replace handpiece on/off valve. |
Air actuator for handpiece is malfunctioning. |
Check to see if kink valve “unkinks” when handpiece is removed from handpiece holder. |
Replace air actuator. |
|
Tubing to handpiece on/off valve and then to air actuator is damaged, loose, or disconnected. |
Check air tubes for damage, looseness, or disconnection. |
Replace damaged tubing. Reconnect any loose or disconnected tubing. |
|
One handpiece does not run, but has water and coolant air. All other handpieces work properly. |
Drive air flow adjustment valve has been adjusted too tightly, closing off air flow to handpiece. |
Check to see if airflow has been cut off. |
Loosen adjustment knob of drive air flow adjustment valve. |
Drive air tubing to handpiece position is damaged, loose, or disconnected. |
Check air tubes for damage, looseness, or disconnection. |
Replace damaged tubing. Reconnect any loose or disconnected tubing. |

Symptom | Probable Cause | Check | Correction |
---|---|---|---|
No water to handpieces or syringe, but handpieces run. |
Water on/off valve is turned off or malfunctioning. |
Check that water on/off valve is turned on. |
Contact licensed plumber to replace on/off valve. |
Master on/off switch valve is turned off or malfunctioning. |
Check to ensure master on/off switch valve is turned on. |
Replace master on/off switch valve if necessary. |
|
City water systems only: Pilot operated valve block malfunctioning or blocked (water side only). |
Check that output pressure is approximately 40 - 50 PSI. |
Clean or flush out pilot operated valve block. If necessary, disassemble and clean pilot operated valve block. |
|
City water systems only: Water regulator setting is too low or malfunctioning. |
Check that water regulator output pressure is 40 PSI. |
Adjust water regulator output pressure until it is 40 PSI. |
|
Self contained water systems only: Air regulator, to water bottle, setting is too low or malfunctioning. |
Check that air regulator output pressure is 40 PSI. |
Adjust air regulator output pressure until it is 40 PSI. |
|
Self contained water systems only: Air tubing to water bottle switch valve or water bottle is damaged, loose, or disconnected. |
Check air tubes for damage, looseness, or disconnection. |
Replace damaged tubing. Reconnect any loose or disconnected tubing. |
|
Self contained water systems only: Water bottle switch valve turned off or malfunctioning. |
Check that water bottle switch valve on side of console is turned on. |
Replace water bottle switch valve if necessary. |
|
Self contained water systems only - Bad water bottle gasket. |
Check for air leaking from top of water bottle. |
Replace water bottle gasket. |

Symptom | Probable Cause | Check | Correction |
---|---|---|---|
All handpieces run and syringe(s) work properly, but there is no water to any of the handpieces. |
Coolant water control valve (anti-retraction valve) is stuck closed. |
Replace suspect coolant water control valve with known working valve. |
Replace coolant water control valve. |
Coolant water toggle valve on foot control is turned off. |
Check that coolant water toggle is turned on. |
Turn coolant water toggle on. |
|
Coolant water toggle valve on foot control is malfunctioning. |
Check for water pressure at output of coolant water toggle valve. |
Replace entire foot control - foot control switch valves are not available. |
|
Orange tubing, signal line from foot control to coolant water control valve, is twisted, pinching off airflow. |
Check if stop pin in arm is damaged or missing. Check if tubing is excessively twisted or pinched. |
Untwist tubing and replace stop pin in arm. |
|
Water tubing from coolant water control valve to kink valves is damaged, loose, or disconnected. |
Check water tube for damage, looseness, or disconnection. Check for any water leakage from tube. |
Replace damaged tubing. Reconnect any loose or disconnected tubing. |
|
Handpiece runs, but there is no water to handpiece. Other handpieces work correctly. |
Coolant water flow adjustment valve has been adjusted too tightly, closing off water flow to handpiece. |
Check to see if water flow has been cut off. |
Loosen adjustment knob of coolant water flow adjustment valve. |
Drive air tubing to handpiece position is damaged, loose, or disconnected. |
Check air tubes for damage, looseness, or disconnection. |
Replace damaged tubing. Reconnect any loose or disconnected tubing. |

Symptom | Probable Cause | Check | Correction |
---|---|---|---|
When removed from handpiece holder, water comes out of any handpiece without foot control pedal being depressed. |
Coolant water toggle valve is malfunctioning. Bleed port is clogged allowing pressure to remain in air line. |
Pull off orange signal line and check for residual air pressure. Residual air pressure indicates bleed port is not working. |
Replace entire foot control - foot control toggle valve is not available. |
Coolant water control valve (anti-retraction valve) stuck open. |
Replace with known working coolant water control valve. |
Replace coolant water control valve. |
|
Coolant water flush valve is malfunctioning or stuck open. |
Replace suspect coolant water flush valve with known working flush valve. |
Replace coolant water flush valve. |
|
One of the coolant water flush check valves is malfunctioning or stuck open. |
Check each coolant water flush check valve to see if it is stuck open. |
Flush out coolant water flush check valve. Replace coolant water flush check valve if necessary. |

Symptom | Probable Cause | Check | Correction |
---|---|---|---|
Water sputters when handpiece is used. |
Air in lines. |
Check for air in coolant water lines. |
Depress coolant water flush valve button and hold for approximately 45 seconds. |
Water is leaking into oil jar. |
Handpiece is not sealing at hose connection point causing water to leak into exhaust hole. |
Check for damaged or missing gasket in handpiece. |
Replace damaged or missing gaskets. |
Water pressure at handpieces and syringes is too low. |
City water systems only: Water regulator setting is too low and needs adjusted. |
Check that water regulator output pressure is 40 PSI. |
Adjust water regulator output pressure until it is 40 PSI. |
Self contained water systems only: Air regulator, to water bottle, setting is too low and needs adjusted. |
Check that air regulator output pressure is 40 PSI. |
Adjust air regulator output pressure until it is 40 PSI. |
|
Self contained water systems only: Air tubing to water bottle switch valve or water bottle is damaged or leaking. |
Check air tubes for damage or looseness. |
Replace damaged tubing. Reconnect any loose tubing. |
|
Self contained water systems only: Water tubing from water bottle to delivery head or syringe(s) is damaged or leaking. |
Check water tubes for damage or looseness. |
Replace damaged tubing. Reconnect any loose tubing. |
|
Self contained water systems only: Water bottle switch valve is partially obstructed. |
Check that water flow at output of water bottle switch is sufficient. |
Replace water bottle switch valve if necessary. |
|
Self contained water systems only: Bad water bottle gasket. |
Check for air leaking from top of water bottle. |
Replace water bottle gasket. |
|
While one handpiece is in use, one or more additional handpieces run even though they are in handpiece holders. |
Suspect handpiece is not stored properly in its handpiece holder, so button of handpiece on/off valve is not being fully depressed. |
Check to see if a kink valve assembly of an unused handpiece is “unkinked”, even though handpiece is in handpiece holder. |
Make sure operators know to properly return handpieces to fully stowed position in handpiece holders. |
Handpiece on/off valve is out of adjustment and is not being activated properly when handpiece is returned to stored position in handpiece holder. |
Check to see if handpiece on/off valve is activated when handpiece is inserted in handpiece holder. |
Adjust handpiece on/off valve button, so it properly activates when handpiece is inserted in handpiece holder. |
|
Handpiece on/off valve on suspect handpiece position is malfunctioning or stuck closed. |
Check to see if a kink valve assembly of an unused handpiece is “unkinked”, even though handpiece is in handpiece holder. Check handpiece on/off valve for proper operation. |
Replace handpiece on/off valve. |
|
Air gauge check valve of handpiece being used is malfunctioning and possibly stuck open allowing drive air to run other handpiece positions. |
Inspect air gauge check valve of handpiece position causing the problem to see if it is stuck open or replace suspect air gauge check valve with known working check valve. |
Replace air gauge check valve. |
|
Leaking tube connector. |
Check for leaks at connection point between handpiece and tube connector. |
Replace tube connector. |
|
Kink valve of suspect position is not kinking properly. |
Check to see if kink valve is leaking. |
Adjust kink valve. Replace kink valve if necessary. |
|
Handpiece runs momentarily after it is removed from the handpiece holder even though foot control pedal is not depressed. |
Drive air foot pedal valve is malfunctioning. Bleed port is clogged allowing pressure to remain in drive air line. |
Pull off drive air line and check for residual air pressure. Residual air pressure indicates bleed port is not working. |
Replace entire foot control- drive air foot pedal valve is not available. |
Main air switch is turned off, but kink valves don’t unkink. |
Drive air foot pedal valve is malfunctioning. Bleed port is clogged allowing pressure to remain in drive air line. |
Pull off drive air line and check for residual air pressure. Residual air pressure indicates bleed port is not working.
|
Replace entire foot control- drive air foot pedal valve is not available. |
Master on/off switch valve malfunctioning or stuck open. |
Replace suspect master on/off switch valve with known working switch valve. |
Replace master on/off valve. |
|
Coolant air only is not working. |
Coolant air flow adjustment valve has been adjusted too tightly, closing off air flow to handpieces. |
Check to see if airflow has been cut off. |
Loosen adjustment knob of coolant air flow adjustment valve. |
Coolant air tubing to handpiece positions is damaged, loose, or disconnected. |
Check air tubes for damage, looseness, or disconnection. |
Replace damaged tubing. Reconnect any loose or disconnected tubing. |

Symptom | Probable Cause | Check | Correction |
---|---|---|---|
Syringe has no air. |
Syringe air flow adjustment valve is closed too tightly. |
Check syringe air flow adjustment valve for proper adjustment. |
Replace syringe air flow adjustment valve if necessary. |
Syringe air tubing to syringe is damaged, loose, or disconnected. |
Check air tubes for damage, looseness, or disconnection. |
Replace damaged tubing. Reconnect any loose or disconnected tubing. |

Symptom | Probable Cause | Check | Correction |
---|---|---|---|
Adjustment screws will not tighten. |
Too much plating. |
- |
Retap adjustment screws and screw holes. |


Symptom | Probable Cause | Check | Correction |
---|---|---|---|
Air lock flex arm will not lock. | Bread tie used for shipping not removed. | Check to see if bread tie is still in place. | Remove bread tie. |
Arm lock cylinder switch valve is malfunctioning. Bleed port is clogged allowing pressure to remain in flex arm cylinder air line. | Pull off flex arm cylinder air line and check for residual air pressure. Residual air pressure indicates bleed port is not working. | Replace arm lock cylinder switch valve. | |
Arm lock cylinder switch valve is malfunctioning or stuck open. | Replace suspect arm lock cylinder switch valve with a known working switch valve. | Replace arm lock cylinder switch valve. |

Symptom | Probable Cause | Check | Correction |
---|---|---|---|
Even though arm lock cylinder is energized, flex arm will not unlock. |
Locking pawls in flex arm have become loose and have slipped off of arm lock cylinder pin. | Check to see if pawls are installed on arm lock cylinder pin. | Re-install locking pawls back on arm lock cylinder pin and tighten mounting hardware. |

Symptom | Probable Cause | Check | Correction |
---|---|---|---|
Air lock flex arm drifts down. | Spring tension jam nut needs adjusted. | - | Tighten jam nut in arm. |
Air lock flex arm drifts up. | Spring tension jam nut needs adjusted. | - | Loosen jam nut in arm. |

Symptom | Probable Cause | Check | Correction |
---|---|---|---|
Air lock flex arm locks but air constantly leaking. | Bad bleed off valve on arm lock cylinder switch valve. | Replace suspect arm lock cylinder switch valve with known working switch valve. | Replace arm lock cylinder switch valve. |

Symptom | Probable Cause | Check | Correction |
---|---|---|---|
- | Too much plating causes adjustment knob to be too difficult to turn. | Check for excessive plating in adjustment knob hole. | Retap hole to clean out excess plating. |


Symptom | Probable Cause | Check | Correction |
---|---|---|---|
Neither cup filler or bowl rinse have water. | Pilot operated valve block malfunctioning. | Replace suspect pilot operated valve block with known working valve block. | Replace pilot operated valve block. |
Main water tubing from pilot operated valve block to cuspidor is damaged, loose, or disconnected. | Check water tubes for damage, looseness, or disconnection. | Replace damaged tubing. Reconnect any loose or disconnected tubing. | |
Water on/off valve is turned off or malfunctioning. | Check that water on/off valve is turned on. | Replace water on/off valve if necessary. |

Symptom | Probable Cause | Check | Correction |
---|---|---|---|
When either cup filler button is pressed, no water comes out. Bowl rinse works correctly. | Cup filler flow adjustment valve has been adjusted too tightly, closing off water to cup filler. | Check to see if water flow to cup filler has been cut off. | Loosen adjustment knob of cup filler flow adjustment valve. |
Tubing between cup filler push button valves and cup filler is damaged, loose, or disconnected. | Check water tubes for damage, looseness, or disconnection. | Replace damaged tubing. Reconnect any loose or disconnected tubing. |

Symptom | Probable Cause | Check | Correction |
---|---|---|---|
When bowl rinse push button is pressed, no water comes out. Cup filler works correctly. | Bowl rinse push button valve is malfunctioning or stuck closed. | Replace suspect bowl rinse push button valve with known working push button valve. | Replace bowl rinse push button valve. |
Water saver control valve is stuck closed. | Replace suspect water saver control valve with known working water saver control valve. | Try to flush any debris out of valve. Replace water saver control valve if necessary. | |
Air pressure from main air regulator is too low and needs adjusted. | Check to see if air pressure from main air regulator is 80 PSI. | Adjust main air regulator output pressure until it is 80 PSI. If necessary, clean or replace air regulator. | |
Tubing between water saver control valve and main air regulator is damaged, loose, or disconnected. | Check air tubing for damage, looseness, or dis-connection. | Replace damaged tubing. Reconnect any loose or disconnected tubing. | |
Adjustable time delay bleed off valve is malfunctioning. | Check to see if air pressure remains in tubing or bleeds off immediately. | Replace adjustable time delay bleed off valve. | |
Bowl rinse flow adjustment has been adjusted too tightly, closing off water to bowl rinse. | Check to see if water flow to bowl rinse has been cut off. | Loosen adjustment knob of bowl rinse flow adjustment valve. | |
Tubing between water saver control valve and bowl rinse tube damaged, loose, or disconnected. | Check water tubes for damage, looseness, or disconnection. | Replace damaged tubing. Reconnect any loose or disconnected tubing. |

Symptom | Probable Cause | Check | Correction |
---|---|---|---|
Water to bowl rinse runs continuously. Cup filler works correctly. | Water saver control valve stuck open. | Replace with known working water saver control valve. Malfunctioning or dirty. | Replace water saver control valve. |
Adjustable time delay bleed off valve is malfunctioning. Bleed port is clogged allowing air pressure to remain in water control saver line after the time delay setting has passed. |
Pull off adjustable time delay bleed off valve line and check for residual air pressure. Residual air pressure indicates bleed port is not working. |
Replace adjustable time delay bleed off valve. |
|
Bowl rinse push button valve is malfunctioning or stuck open. | Replace suspect bowl rinse push button valve with known working valve. | Replace bowl rinse push button valve. |

Symptom | Probable Cause | Check | Correction |
---|---|---|---|
Water to cup filler runs continuously. Bowl rinse works correctly. | One of the cup filler push button valves stuck open. | Check cup filler push button valves or replace suspect cup filler push button valve with known working valve. | Replace cup filler push button valve. |

Symptom | Probable Cause | Check | Correction |
---|---|---|---|
Bowl rinse stops immediately after button is released or does not run long enough (up to 45 seconds) | Adjustable time delay bleed off valve needs adjusted. | Adjust time delay bleed off valve. | Adjust set screw of time delay bleed off valve under cuspidor. |
Adjustable time delay bleed off valve is malfunctioning. Bleed port is open allowing pressure to escape from water control saver line too quickly. | Replace suspect adjustable time delay bleed off valve with known working bleed off valve. | Replace adjustable time delay bleed off valve. |


Symptom | Probable Cause | Check | Correction |
---|---|---|---|
No water is present at syringe and water quick connect. | Air or water on/off valves are turned off or are malfunctioning. | Check that air or water on/off valves are turned on. | Replace air or water on/off valves if necessary. |
Master on/off switch valve is turned off or malfunctioning. | Check to ensure master on/off switch valve is turned on. | Replace master on/off switch valve if necessary. | |
City water systems only: Pilot operated valve block malfunctioning or blocked (water side only). | Check that output pressure is approximately 40 - 50 PSI. | Clean or flush out pilot operated valve block. If necessary, disassemble and clean pilot operated valve block. | |
City water systems only: Water regulator setting is too low or malfunctioning. | Check that water regulator output pressure is 40 PSI. | Adjust water regulator output pressure until it is 40 PSI. If necessary, clean or replace water regulator. | |
Distilled water systems only. Water bottle is empty. | Check that there is water in water bottle. | Fill water bottle with distilled water. | |
Distilled water systems only. Air regulator, to water bottle, setting is too low or malfunctioning. | Check that air regulator output pressure is 40 PSI. | Adjust air regulator output pressure until it is 40 PSI. If necessary, clean or replace air regulator. | |
Tubing between manifold or distilled water bottle and syringe is damaged, loose, or disconnected. | Check water tubes for damage, looseness, or disconnection. | Replace damaged tubing. Reconnect any loose or disconnected tubing. | |
Air or water on/off valves are turned off or are malfunctioning. | Check that air or water on/off valves are turned on. | Replace air or water on/off valves if necessary. | |
Syringe water flow adjustment valve has been adjusted too tightly, closing off water to syringe. | Check to see if water flow to syringe has been cut off. | Loosen adjustment knob of syringe water flow adjustment valve. | |
Distilled water systems only: Water bottle switch valve turned off or malfunctioning. | Check that water bottle switch valve on side of console is turned on. | Replace water bottle switch valve if necessary. | |
Distilled water systems only: Bad water bottle gasket. | Check for air leaking from top of water bottle. | Replace water bottle gasket. |