Procenter Systems and Whip Unit Typical Regulator Removal / Installation Test and Repair
Procenter Systems and Whip Unit Typical Regulator Removal / Installation

1.Turn air and water manual shut-off valves, (1) OFF.
2.Tag and disconnect any tubing connected to fittings of regulator (2).
3.Remove locking ring (3) from regulator (2) then, if needed, remove from mounting plate (4).
If removing water bottle air regulator w/o gauge, the following step is not necessary. Go to step 5.
4.Unscrew regulator (2) from master pilot shut-off valve (5).
5.Remove gauge (6), and any fittings (7) or plugs (8) from regulator (2).

1.Install gauge (6) and any fittings (7) or plugs (8), removed in step 5 of removal procedure, into regulator (2).
If installing water bottle air regulator w/o gauge, the following step is not necessary. Go to step 3.
2.Apply removable Loctite, then screw regulator (2) onto fitting of master pilot shut-off valve (5).
3.Secure regulator (2) with locking ring (3).
4.Connect tubing to proper fittings of regulator.
5.Turn air and water manual shut-off valves (1) ON and check for proper operation.