Elevance® Delivery Handpiece Connection Wiring / Tubing Diagrams
![Closed](../../Skins/Default/Stylesheets/Images/transparent.gif)
![Closed](../../Skins/Default/Stylesheets/Images/transparent.gif)
Model |
Elevance Delivery Systems |
Serial Number | V2220 thru V1736201 |
Handpiece slot position(s) may vary.
![Closed](../../Skins/Default/Stylesheets/Images/transparent.gif)
Model |
Elevance Delivery Systems |
Serial Number | V1736202 to V2216571 |
Handpiece slot position(s) may vary.
![Closed](../../Skins/Default/Stylesheets/Images/transparent.gif)
Model |
Elevance Delivery Systems |
Serial Number | V2216572 to Present |
Handpiece slot position(s) may vary.
![Closed](../../Skins/Default/Stylesheets/Images/transparent.gif)
![Closed](../../Skins/Default/Stylesheets/Images/transparent.gif)
Model |
153800-003 |
Serial Number | V1431117 thru 1506387 |
Handpiece slot position(s) may vary.
![Closed](../../Skins/Default/Stylesheets/Images/transparent.gif)
Model |
153800-003 |
Serial Number | V1506388 thru V1736201 |
Handpiece slot position(s) may vary.
![Closed](../../Skins/Default/Stylesheets/Images/transparent.gif)
Model |
153800-003 |
Serial Number | V1736202 thru V2216571 |
Handpiece slot position(s) may vary.
![Closed](../../Skins/Default/Stylesheets/Images/transparent.gif)
Model |
153800-003 |
Serial Number | V2216572 thru Present |
Handpiece slot position(s) may vary.
![Closed](../../Skins/Default/Stylesheets/Images/transparent.gif)
![Closed](../../Skins/Default/Stylesheets/Images/transparent.gif)
Model |
153800-003 |
Serial Number | V1431117 thru 1506387 |
Pressure sensor tubing permanently connected to pressure sensor. Do not disconnect or damage. When necessary, disconnect pressure sensor tubing from drive air bleed off end.
Test setting, enable, and boost connections by checking voltage while pressing foot pedal. Voltage should decrease to zero.
![Closed](../../Skins/Default/Stylesheets/Images/transparent.gif)
Model |
153800-003 |
Serial Number | V1506388 thru V2216571 |
Pressure sensor tubing permanently connected to pressure sensor. Do not disconnect or damage. When necessary, disconnect pressure sensor tubing from drive air bleed off end.
Test setting, enable, and boost connections by checking voltage while pressing foot pedal. Voltage should decrease to zero.
![Closed](../../Skins/Default/Stylesheets/Images/transparent.gif)
Model |
153800-003 |
Serial Number | V2216572 thru Present |
Pressure sensor tubing permanently connected to pressure sensor. Do not disconnect or damage. When necessary, disconnect pressure sensor tubing from drive air bleed off end.
Test setting, enable, and boost connections by checking voltage while pressing foot pedal. Voltage should decrease to zero.
![Closed](../../Skins/Default/Stylesheets/Images/transparent.gif)
![Closed](../../Skins/Default/Stylesheets/Images/transparent.gif)
Model |
153800-003 |
Serial Number | V1506388 thru V2216571 |
Pressure sensor tubing permanently connected to pressure sensor. Do not disconnect or damage. When necessary, disconnect pressure sensor tubing from drive air bleed off end.
![Closed](../../Skins/Default/Stylesheets/Images/transparent.gif)
Model |
153800-003 |
Serial Number | V2216572 thru Present |
Pressure sensor tubing permanently connected to pressure sensor. Do not disconnect or damage. When necessary, disconnect pressure sensor tubing from drive air bleed off end.
![Closed](../../Skins/Default/Stylesheets/Images/transparent.gif)
Some curing light models connect to the Interface Board with two black wires.
24 VAC to power module and 5 VDC to curing light handpiece are always present.
Curing light is not air driven.
Original Kink Valves Shown.
![Closed](../../Skins/Default/Stylesheets/Images/transparent.gif)
![Closed](../../Skins/Default/Stylesheets/Images/transparent.gif)
Mounted To |
Elevance Dental Chair | Elevance Dental Chair |
Model | Elevance Standard Delivery Unit: 153831-004 | Procenter Delivery Unit: 153831-003 |
A 0 VDC measurement can also indicate connector issue. Check indicator light on Sopro camera docking station.
Make both wire connections in one connector. Each connection can be either color.
In harness "A" (bottom wire group), red and blue wires provide camera power. White wire connection is labeled “luminance - Y” and black wire is labeled “chroma - c”.
Handpiece slot position(s) may vary.
Original Kink Valves Shown.
Item | Part Number | Description |
---|---|---|
A |
002-1526-00 |
Camera Cable Kit |
B |
015-3431-00 | Camera Harness |
Color Abbreviation Chart | ||
NSS = Not Sold Separately | AR = As Required | NLA = No Longer Available | CFN = Call for Number Always Specify Model and Serial Number |
![Closed](../../Skins/Default/Stylesheets/Images/transparent.gif)
Mounted To |
Elevance Dental Chair | Elevance Dental Chair | Elevance Dental Chair |
Model | Elevance Continental-Style Delivery Unit: 153831-009 | Elevance Standard Delivery Unit: 153831-010 | Procenter Delivery Unit: 153831-011 |
Make both wire connections in one connector. Each connection can be either color.
Handpiece slot position(s) may vary.
Original Kink Valves Shown.
![Closed](../../Skins/Default/Stylesheets/Images/transparent.gif)
Mounted To |
Elevance Dental Chair | Elevance Dental Chair | Elevance Dental Chair |
Model | Elevance Continental-Style Delivery Unit: 153831-015 | Elevance Standard Delivery Unit: 153831-016 | Procenter Delivery Unit: 153831-014 |
Original Kink Valves Shown.
![Closed](../../Skins/Default/Stylesheets/Images/transparent.gif)
Mounted To |
Elevance Dental Chair |
Serial Number | All |
For Elevance Continental-Style Delivery Unit, magnet pressed into top cover.
Handpiece slot position(s) may vary.
Original Kink Valves Shown.
![Closed](../../Skins/Default/Stylesheets/Images/transparent.gif)
Handpiece slot position(s) may vary.
Original Kink Valves Shown.
![Closed](../../Skins/Default/Stylesheets/Images/transparent.gif)
![Closed](../../Skins/Default/Stylesheets/Images/transparent.gif)
Model |
153800-003 |
Serial Number | V2200 thru V2216571 |
Syringes can be mounted on either side of the delivery unit as per user's preference.
Standard syringe does not have module or electrical connections. Flow is controlled by rotating knobs and using buttons on syringe.
![Closed](../../Skins/Default/Stylesheets/Images/transparent.gif)
Model |
153800-003 |
Serial Number | V2216572 thru Present |
Syringes can be mounted on either side of the delivery unit as per user's preference.
Standard syringe does not have module or electrical connections. Flow is controlled by rotating knobs and using buttons on syringe.