Whip Unit Master On/Off Switch Valve Removal / Installation Test and Repair
Whip Unit Master On/Off Switch Valve

1.Turn air manual shut-off valve in junction box Off.
2.Press air button of syringe(s) until all residual air pressure is removed from the system.
4.Disconnect the tubing (1) from the Master On/Off toggle valve (2).
5.While holding the threaded ring (3) that secures the On/Off Toggle Valve (2) to the case, loosen the jam nut (4).
6.Remove the threaded ring (3), then remove the Master On/Off Valve (2), jam nut (4) and lockwasher (5).

1.Place the jam nut (4) and lockwasher (5) on to the On/Off Toggle Valve (2) and place it in position in the bottom cover.
Position the outlet (top fitting) on the valve upward to assure the toggle works in conjunction with the On/Off symbols on the cover.
2.Place the threaded ring (3) onto the end of the On/Off Toggle valve and tighten.
3.Connect the 1/8" red tubing to the top (outlet) fitting and the 1/8" purple tubing to the end (inlet) fitting.