Midmark® 1000 Checking for Leaks Test and Repair

Midmark® 1000 Checking for Leaks

Use soapy water to check for leaks.

Refer to: Access Procedures

1.Listen to foot control for air leaks.Tubing inside back of cart.

2.Air regulator and T fittings.

3.Storage air tank and tank fittings.

4.Tri-block and instrument manifolds.

5.Air and water quick disconnects.

6.Handpiece docks.

7.Water bottles.

8.Solenoid muffler, check valve and compressor fittings.