Procenter Cuspidor Bowl Removal / Installation Test and Repair
1.Turn master ON/OFF switch on delivery unit to OFF position.
To prevent residual water spilling, place finger over end of spouts when removing.
2.Remove cup fill spout (1), cup fill gasket (2), bowl flush spout (3), bowl flush gasket (4), and gold trap (5).
Cuspidor bowl is pressure-fit. There is no hardware to remove.
3.Remove cuspidor bowl (6).
1.Position cuspidor bowl (6) over bottom housing (7) and gently press downward.
2.Install cup fill gasket (2) onto cup fill spout (1); then insert cup fill spout into hole (A).
3.Install bowl flush gasket (4) onto bowl flush spout (3); then insert bowl flush spout into hole (B).
4.Install gold trap (5).
5.Turn master ON/OFF switch ON and check for proper operation.