Procenter Cuspidor Troubleshooting

Symptom | Probable Cause | Check | Solution |
---|---|---|---|
Neither cup filler or bowl rinse have water. Handpieces and syringe work properly. | Damaged or disconnected tubing. | Check 1/4” blue tubing, between J-box and cuspidor, for damage and proper connections. | Replace damaged tubing. Reconnect tubing if necessary. |
Check 1/8” tubing, between cuspidor and console (connection box under seat on LR chair), for damage and proper connections. | Replace damaged tubing. Reconnect tubing if necessary. |

Symptom | Probable Cause | Check | Solution |
---|---|---|---|
When either cup filler button is pressed, no water comes out. Bowl rinse works properly. | Cup filler flow adjustment valve set too low. | - | Adjust cup filler flow adjustment knob. |
Cup fill push button valves malfunctioning. | Replace suspect cup fill push button valves with known working valves. | Replace cup fill push button valves. | |
Cup fill (pilot / primary) air valve malfunctioning. | Replace suspect valve with known working valve. | Replace cup fill (pilot / primary) air valve. | |
Cup fill water valve malfunctioning. | Replace cup fill water valve with known working valve. |
Try to flush any debris out of valve. Replace cup fill water valve if necessary. |
|
Damaged or disconnected tubing. | Check 1/8” white tubing, between two cup filler push button valves and shuttle valve, and 1/8” white tubing, between shuttle valve and cup filler air timing valve, for damage and proper connections. | Replace damaged tubing. Reconnect tubing if necessary. | |
Check 1/4” yellow tubing, between two cup filler air timing valves, for damage. | Replace damaged tubing. Reconnect tubing if necessary. | ||
Check 1/8” red tubing, between two cup filler air timing valves and cup fill water valve, for damage. | Replace damaged tubing. Reconnect tubing if necessary. | ||
Check 1/4” blue tubing, between two cup filler adjustment valve and cup fill water valve, for damage. | Replace damaged tubing. Reconnect tubing if necessary. |

Symptom | Probable Cause | Check | Solution |
---|---|---|---|
When bowl flush push button is pressed, no water comes out. Cup filler works properly. | Bowl flush push button valve malfunctioning. | Replace push button valve with known working valve. | Replace bowl flush push button valve. |
Bowl flush control pinch valve adjusted too tightly. | Check adjustment of valve. | Adjust pinch valve. | |
Bowl flush water valve is malfunctioning. | Replace suspect valve with known working valve. |
Try to flush any debris out of valve. Replace bowl flush water valve if necessary. |
|
Damaged or disconnected tubing. | Check all 1/8” red tubing, between bowl flush push button valve, bowl flush solenoid, bowl flush air timing valve, and bowl flush water valve, for damage and proper connections. | Replace damaged tubing. Reconnect tubing if necessary. | |
Check 1/4” blue tubing, between bowl flush pinch valve and bowl flush water valve, for damage and proper connections. | Replace damaged tubing. Reconnect tubing if necessary. |

Symptom | Probable Cause | Check | Solution |
---|---|---|---|
Water to bowl flush runs continuously. Cup filler works properly. | Bowl flush push button valve malfunctioning. | Replace suspect valve with known working valve. | Replace bowl flush push button valve. |
Bowl flush water valve is malfunctioning. | Replace suspect valve with known working valve. |
Try to flush any debris out of valve. Replace bowl flush water valve if necessary. |
|
Bowl flush solenoid is malfunctioning. | Check for voltage across solenoid. |
If no voltage is present, replace bowl flush solenoid. If voltage is present, replace chair PC board. Refer to service information for dental chair. |

Symptom | Probable Cause | Check | Solution |
---|---|---|---|
Water to cup filler runs continuously. Bowl rinse works properly. | One of the cup filler push button valves stuck open. | Replace suspect valve with known working valve. | Replace cup filler push button valve. |

Symptom | Probable Cause | Check | Solution |
---|---|---|---|
Bowl flush stops immediately after button is released or does not run for desired length of time (too long / too short). | Bowl flush timing adjustment valve out of adjustment or malfunctioning. | Adjust bowl flush timing adjustment valve. | Replace bowl flush timing adjustment valve. |

Symptom | Probable Cause | Check | Solution |
---|---|---|---|
Cup fill stops immediately after button is released or does not run for desired length of time (too long / too short). | Cup fill timing adjustment valve is out of adjustment or malfunctioning. | Adjust cup fill timing valve. | Replace cup fill timing adjustment valve. |

Symptom | Probable Cause | Check | Solution |
---|---|---|---|
Cuspidor bowl overflows. | Damaged drain line. | Check drain line for damage. | Replace drain line. |
Clogged drain line or facility drain. | Check drain line and facility drain for clogs. | Clear drain of any possible obstructions. |