Procenter Systems Water Bottle Gasket Replacement Test and Repair
Procenter Systems Water Bottle Gasket Replacement

The illustration above shows a Procenter Console mounted water bottle. The procedure for LR and casework mounted systems is the same.
1.Unscrew water bottle (1) and set aside.
2.Remove damaged gasket (2) from bottle mounting cap (3).

When performing the following step, be sure gasket is properly aligned with top of water bottle. Failure to comply could result in damage to gasket.
1.Align new gasket (2) on top of water bottle (1).
2.Route tubing into water bottle (1); then screw bottle into bottle mounting cap (3)