253 Light Specifications, Dimensions, and Serial Number Location

Equipment is not suitable for use in the presence of a flammable anesthetic mixture with oxygen, air, or nitrous oxide.
Clarification: Equipment is suitable for use in the presence of oxygen, air, or nitrous oxide.
Specifications Chart | |
---|---|
LED | Life: approximately 50,000 hr per bulb manufacturer - Non-replaceable |
Light Pattern | Circular |
Adjustable Spot Size @ Focal distance: 16” (41 cm) | Diameter: 4" (10 cm) to 12" (30 cm) |
Illumination @ Focal distance: 16” (41 cm) Spot size: 4” (10 cm) |
1,700 foot-candles / 18,298 lux |
Color Temperature | 4,500K |
Power Cord Length | 8 ft. (2.5 m) |
Flex Arm Reach from power source enclosure | 54.5” (138 cm) |
Flex Arm Rotation at power source enclosure | 140° |
Electrical Rating | 115 VAC, 50/60 Hz, 100mA |
230 VAC, 50/60 Hz, 50mA | |
Fuses | 1.6 amp, 250V, 5mm x 20mm, Type T low breaking capacity IEC 127-2/3 |
Regulatory Compliance |
UL 60601-1 Rev. 26-Apr-06 CAN/CSA C22.2 No. 60601-1:08 Canadian Medical Device Regulations 5OR/98-282 NFPA99 - Applicable to product requirements |
Classification | Class 1, Type B Applied Part |