354 Light Replacing Mobile 230V Fuses Test and Repair
The voltage setting that appears in the window of the fuse drawer, must match the facility voltage. To adjust the voltage setting, refer to Electrical Input Setting procedure.
Locking tabs are fragile. Pull gently to prevent damage.

1.Turn light switch to STANDBY.
2.Disconnect power cord from light.
3.Using a flat-bladed screwdriver, push in on two locking tabs, then pull fuse drawer out.
4.Pull fuses out of fuse drawer to inspect.
5.Insert fuses into fuse drawer.
6.Insert fuse drawer into AC inlet. Push in until you hear a clicking sound.
7.Connect power cord.

1.Pull out on two selector lock tabs, then pull voltage selector out of fuse drawer.
2.Rotate voltage selector until the correct setting aligns with the voltage window.
3.Insert voltage selector into fuse drawer.