355 Light Bulb Test and Repair

355 Light Bulb Test and Repair

WARNING

The maximum bulb wattage for this light is 100 Watts. There is a risk of fire if this limit is exceeded. Use Midmark part number: 002-0640-00.

Caution

To prevent electrical shock, the light must be in STANDBY mode Standby Mode Symbol. Allow the light to cool before removing bulb. Move the lighthead away from the exam site to prevent contaminants from falling into the area.

Equipment Alert

Halogen bulbs are sensitive to body oils. Do not touch the glass portion of the bulb. If the glass portion of the bulb is handled, wipe with a clean soft lint-free cloth. Wipe with alcohol, then pat dry.

To Replace Bulb

1.Turn light switch to STANDBY Standby Mode Symbol.

2.Loosen three screws on top of lighthead. Do not attempt to remove these screws.

3.Separate top cap from lighthead housing.

4.Pull bulb out of bulb socket, then discard.

5.Grasp new bulb with a clean cloth, then insert into bulb socket.

6.Align three top cap screws with holes, then tighten screws.