M3 Sterilizer Cycle Parameters

The table below shows the temperature / pressure / time parameters for the preset cycles.

Before sterilizing any item in the M3, refer to Loading the Tray in the user's guide.

CYCLE STERILIZATION PARAMETERS DRY TIME ITEMS TO BE STERILIZED  

Unwrapped

Unwrapped

Temperature: 270° F (132° C)

Pressure: 27.1 psi (186 kPa)

Time: 3:30 Minutes

Time: 25 Minutes

Dental Instruments / handpieces loose on a tray.

Other items manufacturers recommend for exposure at 270° F ( 132° C), loose on a tray.

 

Pouches

Pouches

Temperature: 270° F (132° C)

Pressure: 27.1 psi (186 kPa)

Time: 5:30 Minutes

Time: 30 Minutes

Dental Instruments / handpieces in pouches, wrapped, or in a wrapped cassette.

Other items manufacturers recommend for exposure at 270° F ( 132° C), in pouches, wrapped, or in a wrapped cassette.

 

Low Temp

Low Temp

Temperature: 250° F (121° C)

Pressure: 15.0 psi (104 kPa)

Time: 20:00 Minutes

Time: 50 Minutes

Rubber or plastic dental items, dental instruments / handpieces loose on a tray, in pouches, wrapped, or in a wrapped or unwrapped cassette.

Other items manufacturers recommend for exposure at 250° F (121° C), loose on a tray, in pouches, wrapped, or in a wrapped or unwrapped cassette.