M3 Sterilizer Non-Coded Troubleshooting Chart
Display / Symptom |
Cause |
Check |
Action |
---|---|---|---|
Instruments are wet after dry cycle is complete. | Sterilizer not level. | Check if sterilizer is level. | Place sterilizer on a level surface. |
Sterilizer is over loaded. | Check if chamber is overloaded. | Reduce load size. | |
Pouches were placed in chamber with paper side up. | Check if pouches are paper side up. | Place pouches plastic side up. | |
Pouches are stacked on top of each other. | Check if pouches are overlapping or stacked on top of each other. | Pouches may overlap slightly, but items must not be layered. | |
Door tray was removed from chamber and reinserted. | Check if door tray was removed from chamber before completion of the dry cycle. | Instruct operator that once the door tray has been removed from chamber, the dry cycle will be terminated. | |
Steam not being released from chamber. | Check for obstruction in plastic tube connecting unit to external condensing tank | Unkink plastic tube, remove obstruction, or replace. | |
Check copper coil in external condensing tank for blockage or damage | Clear obstruction or replace copper coil. | ||
Check vent port in chamber for blockage. Check filter for blockage | Clear obstruction in vent port. Clean or replace filter. | ||
Check vent valve for blockage. Put unit into Service Diagnostics and check vent valve operation. | Clean or replace vent valve. | ||
Door tray is not opening to its fullest staged extent. | Check if door tray is opening enough to allow steam to escape from chamber. | Adjust door motor throw. | |
Surface heater(s) malfunctioning. | Run Service Diagnostics and check operation of surface heater(s). | Contact Midmark's Technical Service Department. |
Display / Symptom |
Cause |
Check |
Action |
---|---|---|---|
Positive biological indicator. No color change, or incomplete change has occurred on chemical indicator strip. |
Sterilizer overloaded. | Check if chamber is overloaded. | Reduce load size. |
Improper usage. | Check user operating procedure. | Demonstrate proper operation to user. | |
Wrong type of biological indicator or chemical indicator being used for this type of sterilizer. | Check biological or chemical indicator to assure it is for steam sterilization and is rated for the cycle times and temperature being used. | Refer to the Users Guide for proper sterilization monitoring guidelines. | |
Biological indicator or chemical indicator being used is past the expiration date. | Check the indicator and / or packaging to determine what the expiration date is. | Use only biological and chemical indicators that haven't expired. | |
Chemical indicator has come into contact with water in sterilizer. | Check with operator on the positioning of indicator in the chamber. | Demonstrate correct positioning of indicator to user. | |
Biological or chemical indicators are being stored in a damp and / or hot environment. | Check condition of the area where indicators are being stored. | Inform operator to follow storage recommendations of indicator manufacturer. | |
Sterilizer malfunction | Put unit into Service Diagnostics and check for stored error codes, correct temperature, and pressure readings. | Repair as needed based on results of Service Diagnostics test results. |
Display / Symptom |
Cause |
Check |
Action |
---|---|---|---|
PERFORM PERIODIC or MONTHLY MAINTENANCE | Sterilizer software senses that the cycle count has reached 7, 14, or 21 days from initial start-up or monthly maintenance. | Refer to: User Guide for periodic maintenance procedures. | Perform periodic maintenance. |
Sterilizer software senses that the cycle count has reached 28 days from initial start-up or monthly maintenance | Refer to: User Guide for monthly maintenance procedures. | Perform monthly maintenance. |
Display / Symptom |
Cause |
Check |
Action |
---|---|---|---|
Printer does not generate data. | Printer is out of paper. | Lift printer cover and check condition of paper. | Install new roll of paper. User Guide |
Printer cartridge is dry. | Check condition of printer cartridge. | Replace printer cartridge. User Guide | |
Printer wire harness is disconnected. | Check printer wire harness connections on printer and on back of sterilizer. | Secure printer wire harness connections. | |
Printer wire harness has broken or open leads. | Check printer wire harness leads for continuity. | Replace printer wire harness. | |
Printer is malfunctioning. | Check for 5 VDC at TP3 & TP4 on PC board. | If voltage is present, replace printer. | |
PC board is malfunctioning. | Check for 5 VDC at TP3 & TP4 on PC board. | If voltage is not present, replace PC board. |
Display / Symptom |
Cause |
Check |
Action |
---|---|---|---|
Chamber fills with water when cycle is aborted with the door closed. | Unit lost power in heatup or sterilize modes due to a power failure or user turned power off. Chamber filled with water from the reservoir as unit cooled. | Ask questions about recent usage patterns and electrical events (storms, outages, etc.). | Replace check valve on copper tubing between the boiler and the pressure interlock assembly. |
Display / Symptom |
Cause |
Check |
Action |
---|---|---|---|
When in User Diagnosic mode, English / Metric display units can not be changed. | Version 1.1.4 main PCB. | If main PCB version is 1.1.4. | On version 1.1.4 main PCB, flip #7 dip switch ON. |