M9/M11 Sterilizer Fill / Vent Valve Cleaning Test and Repair

WARNING

To prevent risk of shock always disconnect power before removing covers or performing any service procedure.

M9/M11 Sterilizer Valve Cleaning and Repair

To clean or repair the fill / vent valve

1.Drain water from reservoir.

2.Vent Valve only: Disconnect compression fitting.

3.Disconnect wires from valve terminals.

4.Disassemble the valve. Vent valve shown. Fill valve similar.

5.Inspect O-rings, remove any debris.

6.Assemble the valve.

7.Connect wires to valve terminals.

8.Vent Valve only: Connect compression fitting.

9.Refill reservoir with distilled water.

10.Run test cycle / check for leaks.