M9/M11 Sterilizer Temperature Sensor Location / Function Test and Repair

M9/M11 Sterilizer Temperature Sensor

During the Fill Mode

The temperature sensor is not monitored.

During the Heat-Up / Sterilization Modes

The temperature sensor continually monitors the chamber temperature and transmits this information to the PC board.

The PC board turns the heating element ON / OFF based on the readings from the temperature sensor.

During the Vent Mode

The temperature sensor continually monitors the chamber temperature and transmits this information to the PC board.

During the Dry Mode

The temperature sensor continually monitors the chamber temperature and transmits this information to the PC board.

If the temperature exceeds 240°F (115°C), the PC board stops the current flow to the heating element until the temperature drops.