M9/M11 Sterilizer Temperature Sensor Location / Function Test and Repair
During the Fill Mode
The temperature sensor is not monitored.
During the Heat-Up / Sterilization Modes
The temperature sensor continually monitors the chamber temperature and transmits this information to the PC board.
The PC board turns the heating element ON / OFF based on the readings from the temperature sensor.
During the Vent Mode
The temperature sensor continually monitors the chamber temperature and transmits this information to the PC board.
During the Dry Mode
The temperature sensor continually monitors the chamber temperature and transmits this information to the PC board.
If the temperature exceeds 240°F (115°C), the PC board stops the current flow to the heating element until the temperature drops.