M9/M11 Sterilizer -033, -034 and -043 Specifications
For 115 VAC models: Use 104 - 127 VAC, 50/60 Hz alternating current only.
For 230 VAC models: Use 207 - 253 VAC, 50/60 Hz alternating current only.
Failure to do so may result in electric shock to personnel and/or damage to sterilizer.
Water Purity Specification Chart | ||
---|---|---|
AAMI ST-46 (ref.) | ||
Evaporate Residue | ≤ 15 mg/l | ≤ 15 ppm |
Silica | ≤ 2 mg/l | ≤ 2 ppm |
Iron | ≤ 0.2 mg/l | ≤ 0.2 ppm |
Cadmium | ≤ 0.005 mg/l | ≤ 0.005 ppm |
Lead | ≤ 0.05 mg/l | ≤ 0.05 ppm |
Rest of heavy metals, excluding Iron, Cadmium, and Lead 1 | ≤ 0.1 mg/l | ≤ 0.1 ppm |
Chloride | ≤ 3 mg/l | ≤ 3 ppm |
Phosphate | ≤ 0.5 mg/l | ≤ 0.5 ppm |
Conductivity (at 20°C) 2 | ≤ 50 μs/cm | TDS ≤ 25 ppm |
pH | 6.5 to 8 | |
Appearance | Colorless, clean, without sediment | |
Hardness | ≤ 0.1 mmol/l (.585 gr/gal) | ≤ 10 ppm |
Electrical Requirements Chart | |
---|---|
To be properly grounded, unit must be connected to a matching, grounded, dedicated, properly polarized receptacle. | |
115 VAC Unit: |
115 VAC, (+/- 10%) 50/60 Hz, 12 amp Dedicated Supply Circuit*: 120 VAC, 50/60 Hz, 15 amp Max. Power Consumption: 1425 Watts |
230 VAC Unit: |
230 VAC, (+/- 10%) 50/60 Hz, 6.5 amp Dedicated Supply Circuit*: 230 VAC, 50/60 Hz, 10 amp Max. Power Consumption: 1500 Watts |
*Power source must have over-voltage limits less than 1500 watts from mains to ground - Installation Category II in accordance to IEC664 | |
Fuse Ratings: | |
115 VAC | F1 0.25 Amp, 250V, 3 AG Slo-blo, 1/4” x 1 1/4” |
F2 15 Amp, 250V, 3 AB Fast Acting, 1/4” x 1 1/4” | |
230 VAC | F1 0.125 Amp, 250V, Medium-Time Lag, 5 x 20 mm |
F2 8 Amp, Type F 250V, Fast Acting, 5 x 20 mm |
M9 Steam SterilizerSpecifications -033 and -034 | |
---|---|
Physical Dimensions: | |
Overall Length with Plug | 20.38 in. (51.8 cm) |
Overall Width | 15.3 in. (38.9 cm) |
Overall Height with Printer | 15.8 in. (40.1 cm) |
Counter Area | 15.3 in. x 18 in. (38.9 cm x 45.7 cm) |
Chamber | 9 in. diameter x 15 deep (22.9 cm x 38 cm) |
Standard Tray, Large | 7 5/16 in. x 12 in. x 7/8 in. (18.6 cm x 30.5 cm x 2.2 cm) |
Standard Tray, Small | 5 5/8 in. x 12 in. x 7/8 in. (14.3 cm x 30.5 cm x 2.2 cm) |
Weights: | |
Weight with Empty Reservoir | 73 lbs. (33.1 kg) |
Weight with Shipping Carton | 81 lbs (36.7 kg) |
Water Reservoir Capacity | 1.1 Gallons (4.1 liters) to Full Mark usable volume is 0.5 gallons (1.9 liters) |
Water Usage (per cycle) | |
Standard Configuration | 120 to 140 ml (varies depending on the cycle) |
Connected to Direct -to-Drain Thermal Reduction system. (VistaCool TM) | 500 ml |
Chamber Pressure at 270 °F (132 °C) | 27.1 psi. (186.2 kPa) |
Safety Valve Setting | 40 psi (275.8 kPa) |
Heat Emission | 5000 BTU / hr during operation |
Certifications |
ASME Boiler and Pressure Vessel Code, Section VIII, Division 1. Canadian Registration Number available |
UL61010-1, IEC 61010-2-040 CAN/CSA C22.2, No. 61010-1, CSA C22.2, No. 61010-2-040-07 Part 2-040 |
M9 Steam SterilizerSpecifications -043 | |
---|---|
Physical Dimensions: | |
Overall Length with Plug | 20.38 in. (51.8 cm) |
Overall Width | 15.3 in. (38.9 cm) |
Overall Height with Printer | 16.1 in. (40.9 cm) |
Counter Area | 15.3 in. x 18 in. (38.9 cm x 45.7 cm) |
Chamber | 9 in. diameter x 15 deep (22.9 cm x 38 cm) |
Standard Tray, Large | 7 5/16 in. x 12 in. x 7/8 in. (18.6 cm x 30.5 cm x 2.2 cm) |
Standard Tray, Small | 5 5/8 in. x 12 in. x 7/8 in. (14.3 cm x 30.5 cm x 2.2 cm) |
Weights: | |
Weight with Empty Reservoir | 73 lbs. (33.1 kg) |
Weight with Shipping Carton | 81 lbs (36.7 kg) |
Water Reservoir Capacity | 1.1 Gallons (4.1 liters) to Full Mark usable volume is 0.5 gallons (1.9 liters) |
Water Usage (per cycle) | |
Standard Configuration | 120 to 140 ml - Varies depending on the cycle |
Connected to Direct -to-Drain Thermal Reduction system. (VistaCool TM ) | 500 ml |
Chamber Pressure at 270 °F (132 °C) | 27.1 psi. (186.2 kPa) |
Electrical Rating: 115 V models | 115 VAC (+/- 10%), 12 Amp, 50/60 Hz |
Fuse Ratings: | |
115 VAC | F1 0.25 Amp, 250 V, Slo-blo, 1/4” x 1 1/4” |
F2 15 Amp, 250 V, Fast Acting, 1/4” x 1 1/4” | |
Safety Valve Setting | 40 psi (275.8 kPa) |
Heat Emission | 5000 BTU / hr during operation |
Certifications ASME Boiler and Pressure Vessel Code, Section VIII, Division 1. Canadian Registration Number available Midmark is an ISO 13485 and ISO 9001 Certified Company Heat Emission |
UL61010-1, 2nd Edition IEC 61010-2-040:2005, 1st Edition CAN/CSA C22.2, No. 61010-1, 2nd Edition CSA C22.2, No. 61010-2-040-07 Part 2-040, - 1st Edition IEC 60601-1-2 Edition4.0 2014-02 CISPR II:2009, A1:2010 Class A, Group 1 IEC 61000-3-2:2014, Part 3-2 IEC 61000-3-3:2014, Part 3-3 ANSI/AAMI ST55:2010(R)2014 ANSI/AAMI/IEC62304 |
M11 Steam Sterilizer Specifications -033 and -034 | |
---|---|
Physical Dimensions: | |
Overall Length with Plug | 23.8 in. (60.5 cm) |
Overall Width | 17.8 in. (45.2 cm) |
Overall Height with Printer | 17.8 in. (45.2 cm) |
Counter Area | 17.8. x 21 in. (45.2 cm x 53.3 cm) |
Chamber | 11 in. diameter x 18 in. deep (28 cm x 46 cm) |
Standard Tray, Large | 9 in. x 15 in. x 1 1/8 in. (22.9 cm x 38 cm x 2.9 cm) |
Standard Tray, Small | 6 5/8 in. x 15 in. x 1 1/8 in. (16.8 cm x 38 cm x 2.9 cm) |
Weights: | |
Weight with Empty Reservoir | 99 lbs. (44.9 kg) |
Weight with Shipping Carton | 131 lbs (59.4 kg) |
Water Reservoir Capacity | 1.4 Gallons (5.3 liters) to Full Mark usable volume is 1.0 gallons (3.8 liters) |
Water Usage (per cycle) | |
Standard Configuration | 120 to 180 ml - Varies depending on the cycle |
Connected to Direct -to-Drain Thermal Reduction system. (VistaCool TM ) | 650 ml |
Chamber Pressure at 270 °F (132 °C) | 27.1 psi. (186.2 kPa) |
Safety Valve Setting | 40 psi (275.8 kPa) |
Heat Emission | 5000 BTU / hr during operation |
Certifications ASME Boiler and Pressure Vessel Code, Section VIII, Division 1. Canadian Registration Number available |
UL61010-1, IEC 61010-2-040 CAN/CSA C22.2, No. 61010-1, CSA C22.2, No. 61010-2-040-07 Part 2-040,
|
M11 Steam Sterilizer Specifications -043 | |
---|---|
Physical Dimensions: | |
Overall Length with Plug | 23.8 in. (60.5 cm) |
Overall Width | 17.8 in. (45.2 cm) |
Overall Height with Printer | 18.1 in. (50.0 cm) |
Counter Area | 17.8. x 21 in. (45.2 cm x 53.3 cm) |
Chamber | 11 in. diameter x 18 in. deep (28 cm x 46 cm) |
Standard Tray, Large | 9 in. x 15 in. x 1 1/8 in. (22.9 cm x 38 cm x 2.9 cm) |
Standard Tray, Small | 6 5/8 in. x 15 in. x 1 1/8 in. (16.8 cm x 38 cm x 2.9 cm) |
Weights: | |
Weight with Empty Reservoir | 99 lbs. (44.9 kg) |
Weight with Shipping Carton | 131 lbs (59.4 kg) |
Water Reservoir Capacity | 1.4 Gallons (5.3 liters) to Full Mark usable volume is 1.0 gallons (3.8 liters) |
Water Usage (per cycle) | |
Standard Configuration | 120 to 180 ml - Varies depending on the cycle |
Connected to Direct -to-Drain Thermal Reduction system. (VistaCool TM ) | 650 ml |
Chamber Pressure at 270 °F (132 °C) | 27.1 psi. (186.2 kPa) |
Electrical Rating: 115 V models | 115 VAC (+/- 10%), 12 Amp, 50/60 Hz |
Fuse Ratings: | |
115 VAC | F1 0.25 Amp, 250 V, Slo-blo, 1/4” x 1 1/4” |
F2 15 Amp, 250 V, Fast Acting, 1/4” x 1 1/4” | |
Safety Valve Setting | 40 psi (275.8 kPa) |
Heat Emission | 5000 BTU / hr during operation |
Certifications ASME Boiler and Pressure Vessel Code, Section VIII, Division 1. Canadian Registration Number available Midmark is an ISO 13485 and ISO 9001 Certified Company Heat Emission |
UL61010-1, 2nd Edition IEC 61010-2-040:2005, 1st Edition CAN/CSA C22.2, No. 61010-1, 2nd Edition CSA C22.2, No. 61010-2-040-07 Part 2-040, - 1st Edition IEC 60601-1-2 Edition4.0 2014-02 CISPR II:2009, A1:2010 Class A, Group 1 IEC 61000-3-2:2014, Part 3-2 IEC 61000-3-3:2014, Part 3-3 ANSI/AAMI ST55:2010(R)2014 ANSI/AAMI/IEC62304 |