M9/M11 Sterilizer Door Switch Replacement Test and Repair
To prevent risk of shock always disconnect power before removing covers or performing any service procedure.
Model |
M9M11 -020 thru -022, -033, -034 |
Serial Number | All |
To replace the door switch
1.Disconnect door switch wires from J1 and J2 on PC board.
2.Remove two screws, nuts, lockwashers, and door switch.
3.Position new door switch in housing.
4.Secure switch with screws, lockwashers, and nuts.
5.Connect door switch wires to J1 and J2 on PC board.
Model |
M9M11 -040 thru -043 |
Serial Number | All |
To replace the door switch
1.Disconnect door switch wires from J1 and J2 on PC board.
2.Remove two screws, nuts, lockwashers, and door switch.
3.Position new door switch in housing.
4.Secure switch with screws, lockwashers, and nuts.
5.Connect door switch wires to J1 and J2 on PC board.