M9/M11 Sterilizer Temperature Sensor Test and Repair
Some procedures require power to be connected with covers removed. Line voltage is present. Use extreme caution to prevent electric shock.
Model |
M9M11 -020 thru -022, -033, -034 |
Serial Number |
All |
1.Allow sensor to cool to room temperature before testing.
2.Disconnect sensor harness from J12 on PC board.
3.Set meter to 2K Ω.
4.Place meter probes on red and white sensor wires as shown.
Meter Reading | Required Action |
---|---|
Acceptable range: 1.07 to 1.1 kΩ |
Perform Supply Voltage Test |
Any reading out of the acceptable range |
Replace temperature sensor |
1.Disconnect sensor harness from J12 on PC board.
2.Set meter to 20 VDC.
3.At J12 on PC board, place meter probes on pins 1 and 2 as shown.
Meter Reading | Required Action |
---|---|
Acceptable range: 4.0 to 6.0 VDC |
Main PC board is OK |
Any reading out of the acceptable range |
Replace Main PC board |
Model |
M9/M9D/M11 -040 thru -043 |
Serial Number | All |
1.Allow sensor to cool to room temperature before testing.
2.Disconnect sensor harness from J12 on PC board.
3.Set meter to 2K Ω.
4.Place meter probes on red and white sensor wires as shown.
Meter Reading | Required Action |
---|---|
Acceptable range: 1.07 to 1.1 kΩ |
Perform Supply Voltage Test |
Any reading out of the acceptable range |
Replace temperature sensor |
1.Disconnect sensor harness from J12 on PC board.
2.Set meter to 20 VDC.
3.At J12 on PC board, place meter probes on pins 1 and 2 as shown.
Meter Reading | Required Action |
---|---|
Acceptable range: 4.0 to 6.0 VDC |
Main PC board is OK |
Any reading out of the acceptable range |
Replace Main PC board |
1.Disconnect sensor harness from J12 on PC board.
2.Set meter to 20 VDC.
3.At J12 on PC board, place meter probes on pins 1 and 2 as shown.
Meter Reading | Required Action |
---|---|
Acceptable range: 3.0 to 4.0 VDC |
Main PC board is OK |
Any reading out of the acceptable range |
Replace Main PC board |
1.Disconnect sensor harness from J12 on PC board.
2.Set meter to 20 VDC.
3.At J12 on PC board, place meter probes on pins 1 and 2 as shown.
Meter Reading | Required Action |
---|---|
Acceptable range: 3.0 to 4.0 VDC |
Main PC board is OK |
Any reading out of the acceptable range |
Replace Main PC board |
1.Disconnect sensor harness from J12 on PC board.
2.Set meter to 20 VDC.
3.At J12 on PC board, place meter probes on pins 1 and 2 as shown.
Meter Reading | Required Action |
---|---|
Acceptable range: 3.0 to 4.0 VDC |
Main PC board is OK |
Any reading out of the acceptable range |
Replace Main PC board |