Midmark M9® -05x / M11® -05x Sterilizer Vent/Fill Valve Test and Repair
To prevent risk of shock always disconnect power before removing covers or performing any service procedure.
Some procedures require power to be connected with covers removed. Line voltage is present. Use extreme caution to prevent electric shock.
Model |
M9 -050, -053 | M11 -050, -051, -053 |
Serial Number | All | All |

1.Set meter to VDC.
2.Place meter probes on desired points and check reading.
• Fill = TP74 and TP75
• Vent = TP72 and TP73
3.Energize the desired valve and check reading. The corresponding LED should light up.
• Fill = LED D6 on
• Vent = LED D8 on
Meter Reading |
Required Action |
---|---|
Energized |
|
Line voltage 115 V model = approximately 108 VDC Line voltage 130 V model = approximately 216 VDC |
PC board is OK. Perform Continuity Test. |
Any other reading |
Replace PC board. |
Not Energized |
|
Approximately 0 VDC |
PC board is OK. |
Any other reading |
Replace PC board. |

1.Set meter to Ohms (Ω).
2.Disconnect wires from desired valve.
3.Place meter probes on valve terminals.
Meter Reading |
Required Action |
---|---|
Approximately k Ωs (Ohms) reading |
Valve is OK. |
OL |
Replace valve. |
Approximately 0 Ωs (Ohms) | Replace valve. |