Midmark M9® -05x / M11® -05x Sterilizer Water Level Sensor Test and Repair
To prevent risk of shock always disconnect power before removing covers or performing any service procedure.
Model |
M9 -050, -053 | M11 -050, -051, -053 |
Serial Number | All | All |
![Closed](../../Skins/Default/Stylesheets/Images/transparent.gif)
Some procedures require power to be connected with covers removed. Line voltage is present. Use extreme caution to prevent electric shock.
Voltage Test
1.Set meter to VDC.
2.Place meter probes on points TP177 and TP292 (as shown in the figure above).
3.Activate the water level sensor by filling chamber with water up to the sensor and check reading.
LED D52 should light up.
Meter Reading |
Required Action |
---|---|
No water touching sensor = approximately 3.3 VDC Water touching sensor (chamber full) = approximately 0 VDC |
PC board is OK . Perform Continuity Test. |
Any other reading |
Replace PC board. |
![Closed](../../Skins/Default/Stylesheets/Images/transparent.gif)
Sterilizer or components may be hot or become hot. Wait until the sterilizer or components are cool to the touch before continuing service.
Resistance Test
1.Empty any water from chamber and dry sensor.
2.Disconnect wire from sensor terminal.
3.Set meter to Ohms (Ω).
4.Place one meter probe on sensor terminal and the other probe on the chamber wall (as shown in the figure above).
Residue buildup on the sensor disk can cause malfunctions during the Filling mode. Clean the sensor disk with Speed-Clean and an abrasive pad; then dry disk thoroughly.
Meter Reading |
Required Action |
---|---|
OL |
Water level sensor is good. |
Any reading other than OL |
Replace water level sensor. |