204 and 604 Table Receptacle System Test and Repair
Model |
204 -001 thru -004 / -006 / -007 / -011 / -012 / -014 | 604 -001 thru -003 / -005 / -006 |
Serial Number | All | All |

There are two receptacle circuit breakers to protect the system. If the receptacle's maximum load is exceeded, the circuit breakers interrupt power to the receptacle.
Maximum Load, both outlets combined = 5 amps
1.To reset circuit breaker(s), push toggle switch inward.

1.Remove receptacle cover.
2.Set meter to read VAC.
3.Place meter probes on wire connection screws of receptacle.
Supply Voltage Test Readings:
If line voltage reading is approximately 115 / 230 VAC, replace receptacle.
If there is no voltage present, check all wire connections from receptacle to supply source.