244 Table Maintenance Schedule

Bariatric Power Treatment Table Maintenance Schedule
Interval Inspection or Service Service, Adjust, Repair, and / or Replace as Required 244
Weekly Cleaning Clean upholstery with appropriate diluted bleach solution 10:1 (water: bleach) X
Wipe painted metal and plastic surfaces with a clean soft cloth and mild cleaner. X
Obvious Damage Visually inspect components for damage that could cause problems during operation or unsafe operation. X
Semi-Annually Mechanical Operation Check all mechanical functions using the hand control. Repeat using the foot control when pressed. X
Table should move smoothly and quietly when any function is activated. X

Check base cover limit switches for proper operation.
A. Press down firmly on the base cover.
B. Activate any table function with the hand / foot control.
If table function does not move, limit switches OK. If table function does move, inspect / test limit switches.

X
Labels/Decals Replace any missing or illegible labels. X
Hardware All fasteners must be present and fastened securely. X
Electrical System Inspect power cord and all wiring for damage. X
Be sure all electrical connections are tight. X

Date of Service:

 

Model Number:

 

Location:

 

Serial Number:

 

Service Technician:

 

Notes: