625 Table Calibration Procedure Test and Repair
When S2 switch is pressed you will hear a steady pattern of beeps. When calibration is completed you will hear three short beeps.
To calibrate
1.Unplug and re-plug power cord.
2.Wait for the power up beep sequence from table, then within two seconds, press S2 switch.
3.Run Base all the way down.
4.Run Base all the way up.
5.Run Back all the way down.
6.Run Back all the way up.