625 Table Scale Calibration Procedure Test and Repair

Model

625 -004
Serial Number All

If abort message is displayed, unplug table for 60 seconds. Remove and reinstall hand control batteries, then repeat step 1.

625 TableScale Calibration Procedure, 625-004 only

1.Remove any load from the patient surface.

2.Press and hold the Weight and Plus buttons until “Pause” is displayed for at least one second, then release buttons.

625 Table Scale Calibration Procedure, 625-004 only

3.When “CLrAL” is displayed, firmly press and release the plus button.

4.After count down, “Ad50” will be displayed.

If something other then “Ad50” is displayed, press the weight button until “Ad50” is displayed. Early units may display “Add 50”.

625 Table Scale Calibration Procedure, 625-004 only

5.Place fifty pound calibrated weight on patient surface.

6.Firmly press and release the plus button.

After count down “donE” will be displayed.

625 Table Scale Calibration Procedure, 625-004 only

7.Wait until display clears, then firmly press and release the weight button.

“50 lbs” should be displayed.

8.Remove calibrated weight.