640 Doran Scale Exam Table Checking / Adjusting Arm Alignment on Linear Scale Test and Repair
To Check Measuring Arm Alignment
1.Pivot measuring arm out as shown.
2.Using a carpenter’s square, check to see if the measuring arm is square with the back of the table top.
3.If necessary, adjust arm until square.
To Adjust Measuring Arm
Using a 3/32” allen wrench, adjust set screw in / out until measuring arm is square with the back of the table top.