640 Tanita Scale Exam Table Linear Scale Test and Repair

640 Exam Table Linear Scale

To check measuring arm alignment

1. Pivot measuring arm out as shown.

2.Using a carpenter’s square, check to see if the measuring arm is square with the back of the table top.

3.If necessary, adjust arm until square.

To adjust measuring arm

1.Using a 3/32” allen wrench, adjust setscrew in/out until measuring arm is square with the back of the table top.