Canis Major Limit Switch Test and Repair
Model |
CMDT | CMDT-WS 45 and 60 |
Serial Number | All | All |
The limit switch is located on the inside of the column and prevents the work surface from lowering beyond a certain point. The control box reduces the voltage supplied to the switch to 24 VDC.
1.Raise Table to maximum height.
2.Remove back panel from column. Refer to: Back Panel
3.Remove bottom nut on limit switch bracket.
4.Swing bracket forward to adjust limit switch. Tighten nut.
5.Lower table to verify limit switch adjusted height.
6.Adjust again if necessary.
7.Install back panel.