QuickClean™ Ultrasonic Cleaners Power Switch Test and Repair

QuickClean™ Ultrasonic Cleaners Power Switch Test and Repair

WARNING

To prevent risk of shock always disconnect power before removing covers or performing any service procedure.

To Replace Power Switch

1.Turn power off, unplug, and drain water from tank.

2.Remove Bottom Plate Assembly. Refer to: Access Procedures

3.Remove the brown and blue wires from the back of the switch.

4.Press in on the side tabs and push the switch out of the case.

5.Replace switch, install wires, and reassemble.