641 Chair Specifications

641 Specifications

641 Specifications 641 Specifications 641 Specifications

Equipment not suitable for use in the presence of a flammable anesthetic mixture with air, or with oxygen, or nitrous oxide.

For packaged weight, contact Midmark Customer Service

Specifications

Patient Weight - maximum

450 lbs - 204 kgs

Range of Motion

See illustration

Dimensions

See illustration

Weight of Table:

w/ Standard Base

440 lbs - 200 kg

w/ Rotational Base

525 lbs - 238 kg

Power Cord

8 ft. - 243.8 cm long

Electrical Requirements:

See Compliance Charts

Foot / Hand Control Voltage

10 VAC, SELV - Safety Extra Low Voltage

Duplex Receptacle - optional maximum load

115 VAC, 5A, 50/60 Hz

Fuses:

Main Power - located at power cord inlet

 

F1 and F2

6.3A, 250V, Type-T, 5 x 20 mm High-Breaking Current

Duty Cycle

Actuator motor run time:

Intermittent Operation - 30 seconds On / 5 minutes off

Classifications

Class 1, Type B Applied Part,

Protection against ingress of fluids

Ordinary Equipment

Foot Control only

PX1

Regulatory Compliance:

See Compliance Charts

Table Speeds - by function

Back / Tilt:

w/Patient Load <200lbs - <90 kg

less than 16 seconds

w/Patient Load >200lbs - >90 kg)

less than 20 seconds

Base

less than 20 seconds

Foot

less than 14 seconds