Midmark® Smart G-Vacuum Fan Test and Repair
230 VAC is supplied to the continuous run fans.
When working on Twin units, unplug parallel harness.
1.Turn power off at on/off switch and main power supply box.
2.Remove rear panel. Refer to: Back Cover
3.Set meter to Ω.
4.Unplug outside fan harness from base unit and fans.
5.Check continuity in fan harness (outside).
6.Unplug inside fan harness from plug on side of electrical box.
7.Check continuity in fan harness (inside). Refer to: Wire Diagrams
| Check | Meter Reading | Required Action |
|---|---|---|
| Continuity on Fan Harness - outside electrical box | < 1 ohm | Fan Harness OK. |
| OL | Replace Fan Harness | |
| Continuity on Fan Harness - inside electrical box | < 1 ohm | Fan Harness OK. |
| OL | Replace Fan Harness |