AH9000 Mobile Delivery Unit Checking for Leaks Test and Repair
Use soapy water to check for leaks.
Refer to: Access Procedures
1.Listen to foot control for air leaks. Tubing inside dental delivery.
2.Air regulator and T fittings.
3.Storage air tank and tank fittings.
4.Tri-block or quadblock and instrument manifolds.
5.Handpiece holders.
6.Water bottles.
7.Solenoid muffler, check valve and compressor fittings.