Elevance® Delivery Instrument Connection General Setup Wiring / Tubing Diagrams
*Refer to: J-Box Connections
Foot control drive air for handpieces must be ON for handpiece functions except syringe, curing light and cameras.
Original Kink Valves Shown.
PSI | kpa |
---|---|
30 | 206.8 |
50 | 344.7 |
80 | 551.6 |
100 | 689.5 |
Model |
153800-003 |
Serial Number | V2200 thru V1736201 |
Selected kink valve connections shown for clarity.
Model |
153800-003 |
Serial Number | V1736202 thru V2216571 |
Pilot Air Key | |
---|---|
Tubing Color | Supplies Air To |
Orange | Water Pilot Valve |
Yellow | Supply Air |
Black | Brake Pilot Valve |
Model |
153800-003 |
Serial Number | V2216572 thru Present |
Pilot Air Key | |
---|---|
Tubing Color | Supplies Air To |
Orange | Water Pilot Valve |
Yellow | Supply Air |
Black | Brake Pilot Valve |
Model |
153800-003 |
Serial Number | V2200 thru V2216571 |
Bypass controls provide a method to manually supply air and water to all the air driven handpieces if the Integration Control Module® loses power. Manual control is available when the air and water bypass toggles are flipped into the BYPASS positions.
If the ICM is powered ON AND the water toggle is in the BYPASS position, the water button override remains an active control.
Model |
153800-003 |
Serial Number | V2216572 thru Present |
Bypass controls provide a method to manually supply air and water to all the air driven handpieces if the Integration Control Module® loses power. Manual control is available when the air and water bypass toggles are flipped into the BYPASS positions.
If the ICM is powered ON AND the water toggle is in the BYPASS position, the water button override remains an active control.